i1Pro TestsGeneral Information
Date and Time: 9/5/2022 12:00:14 PM
Application version: 4.2.0.347, 32-bit
SDK version i1Pro: 4.2.4.5061
SDK version i1iO: 2.2.4.5061
SDK version i1Pro3: 1.0.5.5806
SDK version i1iO3: 1.0.4.5806
SDK version i1iSis: 2.0.0.4968
SDK version ColorMunki: 1.0.3.4630
SDK version i1Display2: X-Rite, Inc. SipCal Library Ver. 7.3.7 (32 bit)
Platform: Microsoft Windows 10, 64-bit
Open instrument
Connection: Successful
Open instrument: Successful
Instrument information
Instrument inspection: Successful
Serial number i1Pro: 846462
Type: i1Pro (M2 only)
Revision: C
Firmware version: 406
CPLD version: 999
Dark measurement count: 7594
Lamp burning time: 26753.3 seconds
Spot measurement count: 2427
Scan measurement count: 5149
Factory calibration: 2/28/2006
Instrument Licenses
Measurement Modes: [Reflectance Spot, Reflectance Scan, Emission]
Licensed functionality: [Has UVcut filter, Spectral values, Tristimulus, Density]
Licensed Modules:
i1Match Module "Scanner": NOT licensed
i1Match Module "Monitor": NOT licensed
i1Match Module "Printer": NOT licensed
i1Match Module "Printer RGB only": NOT licensed
i1Match Module "Printer CMYK only": NOT licensed
i1Match Module "Beamer": NOT licensed
i1Match Module "Printer Easy RGB": NOT licensed
i1Match Module "Printer Easy CMYK": NOT licensed
i1Match Module "Digicam": NOT licensed
i1Match Module "Editor": NOT licensed
i1Profiler Module "Display Profiling": NOT licensed
i1Profiler Module "RGB Printer Profiling": NOT licensed
i1Profiler Module "CMYK Printer Profiling": NOT licensed
i1Profiler Module "Printer QA": NOT licensed
i1Profiler Module "Advanced-Mode": NOT licensed
i1Profiler Module "Projector": NOT licensed
Button test
Button: Successful
Emission test
Emission calibration: Successful
Prepare the instrument for emission measurement
White measurement: Successful
Gray measurement: Successful
Black measurement: Successful
Red measurement: Successful
Green measurement: Successful
Blue measurement: Successful
Emission: Successful
Button test
Button: Successful
Noise Tests
Dark measurement: Successful
Noise measurements: Successful
Noise result: Successful
Reflectance test
Reflectance Spot calibration: Successful
Reflectance Spot measurement 1: Successful
Reflectance Spot measurement 2: Successful
Reflectance Spot measurement 3: Successful
Reflectance Spot measurement 4: Successful
Reflectance Spot measurement 5: Successful
Reflectance Spot measurement 6: Successful
Reflectance Spot measurement 7: Successful
Reflectance Spot measurement 8: Successful
Reflectance Spot measurement 9: Successful
Reflectance Spot measurement 10: Successful
Lamp test: Successful
Reflectance drift test: Successful
Reflectance Scan calibration: Successful
*** PASS ***